Title: Comparison of Diagon Coag XL and Cobas t 511 analyzers
Authors: Guven, Berrak
Benice, Ismail
Can, Murat
Keywords: Analytical performance
Cobas t 511
Diagon Coag XL
Issue Date: 2022
Publisher: International Journal of Medical Biochemistry
Series/Report no.: Technical Report;116-119
Abstract: Objectives: We compared prothrombin time (PT), activated partial thromboplastin time (aPTT), fibrinogen, and Ddimer test results measured using the Diagon Coag XL coagulation analyzer with Cobas t 511 analyzer. Methods: Imprecision studies were performed for the PT, aPTT, fibrinogen, and D-dimer tests used by the Diagon Coag XL analyzer. For the comparison study, we used the leftover citrated plasma from patient samples after routine analysis with Cobas t 511. All of the results were analyzed using the correlation coefficient and Passing–Bablok regression analysis. Results: Total coefficient of variation obtained for all tests were within the criteria for acceptance. The method comparison study showed a good correlation between the results obtained on Diagon Coag XL and Cobas t 511 analyzers, except for aPTT test. The correlation coefficients obtained were 0.98 for PT, INR, and D-dimer, 0.95 for fibrinogen, and 0.80 for aPTT. Conclusion: For PT, aPTT, fibrinogen, and D-dimer tests, Diagon Coag XL analyzer is suitable for monitoring the coagulation system, and it can be used in clinical laboratories. However, the precision values of tests stated by the manufacturer must be verified.
URI: http://localhost:8080/xmlui/handle/123456789/8149
ISSN: 2618-642X
Appears in Collections:Vol 5 No 2 (2022)

Files in This Item:
File Description SizeFormat 
116 - 119.pdf657.95 kBAdobe PDFView/Open


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.